Dako Hybridizer | Handbook 1 Hybridizer | Handbook
10 Dako Hybridizer | Handbook Section 2 | System Overview Principle and Intended Use For in vitro diagnostic use to co-denature and hybridize
Dako Hybridizer | Handbook 11 Symbols and Definitions (cont.) Display Abbreviations PGM Program Denat & Hyb Denaturation & Hybridization
12 Dako Hybridizer | Handbook Audible Indicators Single beep – on all legal keystrokes. Two quick beeps – upon accepting a field and cursor has mov
Dako Hybridizer | Handbook 13 High ambient temperature condition: The instrument will attempt to achieve process set temperatures. However, if
14 Dako Hybridizer | Handbook If the ambient temperature changes during a process and causes the instrument process set temperature to change beyon
Dako Hybridizer | Handbook 15 If the ambient temperature changes after a process is completed, but before the user removes the slides and causes
16 Dako Hybridizer | Handbook Section 3 | Programming Opening and Closing the Lid The plate may be hot. Use caution and check temperature on di
Dako Hybridizer | Handbook 17 Note: If all 40 program numbers have been used, the “Create a PGM” line in the Main Menu will no longer appear. An
18 Dako Hybridizer | Handbook Create a Hybridization Only Program (Hyb Only) Turn unit on and wait for the Main Menu screen. Use the arrows keys to
Dako Hybridizer | Handbook 19 Create a Fixed Temperature Program (Fixed Temp) Use the arrows keys to move cursor to “Create a PGM” line and pres
2 Dako Hybridizer | Handbook Hybridizer | Handbook Document Number 0003061 Revision C July 2007 For In Vitro Diagnostic Use
20 Dako Hybridizer | Handbook Section 4 | Operating Instructions Opening and Closing the Lid The plate may be hot. Use caution and check temp
Dako Hybridizer | Handbook 21 Display will confirm PGM number/name and Denat & Hyb times and temperatures. Cursor highlights “Run PGM” line.
22 Dako Hybridizer | Handbook The Hybridizer will automatically cool to hybridization set temperature once denaturation is completed. “Ambient
Dako Hybridizer | Handbook 23 Display will confirm PGM number/name and Hyb Only times and temperatures. Cursor highlights the “Run PGM” line.
24 Dako Hybridizer | Handbook Upon program completion Hybridizer will beep five times and the display will show “Process Complete”. Hybridization t
Dako Hybridizer | Handbook 25 When the fixed temp set temperature is reached the display prompts to “Add Slides and Close Lid”. Before adding sli
26 Dako Hybridizer | Handbook Recommended for Best Results Place glass coverslip over probe and sample Apply sealant along all coverslip edge
Dako Hybridizer | Handbook 27 Predefined Limits Program Mode Temperature Range Timer Limits Denature 50°C to 99°C 0-30 minutes Hybridization Ro
28 Dako Hybridizer | Handbook Section 5 | Edit a Program Opening and Closing the Lid The plate may be hot. Use caution and check temperature
Dako Hybridizer | Handbook 29 Cursor highlights existing program type: “Denat & Hyb”, “Hyb only” or “Fixed Temp”. Press “Enter” button to acc
Dako Hybridizer | Handbook 3 Copyright© 2007 Dako Colorado, Inc. All rights reserved This document may not be copied in whole or in part or re
30 Dako Hybridizer | Handbook Section 6 | Maintenance Overview Dako recommends that instrument operators perform periodic inspections and preve
Dako Hybridizer | Handbook 31 Troubleshooting Unit does not turn on, no power. Check both cord ends are plugged in. Check Fuses located on rear
32 Dako Hybridizer | Handbook Appendix A | Specifications Code Numbers S 2450 S 2451 Electrical 100-120 VAC @ 3.0 A 220-240 VAC @ 1.6A P
4 Dako Hybridizer | Handbook Table of Contents | Handbook Section 1 | Unpacking and Installation...
Dako Hybridizer | Handbook 5 Section 5 | Edit a Program... 28 O
6 Dako Hybridizer | Handbook How to use this handbook This handbook along with information contained on product labels should provide all the inf
Dako Hybridizer | Handbook 7 This page is intentionally blank.
8 Dako Hybridizer | Handbook Section 1 | Unpacking and Installation Inspect Packaging The Hybridizer and its accessories are delivered in one c
Dako Hybridizer | Handbook 9 This page is intentionally blank.
Komentarze do niniejszej Instrukcji